- Insulation Resistance up to 10TΩ
- Short-Circuit Current up to 5mA
- Wide Test Voltage from 250V to 5000V
- Diagnostic Insulation Tests: IR, PI, DAR, DD, SV, RAMP (Breakdown and Burn modes), Capacitance, Leakage Current
- Wireless communication by Bluetooth for transferring and showing real-time data to PC and Android device
- Memory and Logging functions
- Filter function reduces noise interference for obtaining stable measurements
- Large back light display with Bar-Graph indication
- Robust design for field use with IP65 (lid closed)
- Powered by rechargeable battery
- Safety Standard IEC 61010-1 CAT Ⅳ 600V
| Insulation resistance | ||||||||
|---|---|---|---|---|---|---|---|---|
| Test voltage | 250V (*1) | 500V | 1000V | 2500V | 5000V | |||
| Max measurement value | 9.99GΩ | 99.9GΩ | 199GΩ | 999GΩ | 9.99TΩ | |||
| Accuracy | 0.0 - 99.9MΩ ±5%rdg±3dgt | 0.0 - 999MΩ ±5%rdg±3dgt | 0.0 - 1.99GΩ ±5%rdg±3dgt | 0.0 - 99.9GΩ ±5%rdg±3dgt | 0.0 - 99.9GΩ ±5%rdg±3dgt | |||
| 0.1G - 9.99GΩ ±20%rdg | 1G - 99.9GΩ ±20%rdg | 2G - 199GΩ ±20%rdg | 100G - 999GΩ ±20%rdg | 0.1T - 9.99TΩ ±20%rdg | ||||
| Short circuit current | Max 5.0mA | |||||||
| Output voltage | Accuracy | -10 - +10% | -10 - +20% | 0 - +20% | ||||
| Variable | - | -20% - 0% (5% step) | ||||||
| Monitor | ±10%rdg±20V | |||||||
| Voltage measurement | ||||||||
| Measuring range | AC: 30 - 600V (50/60Hz), DC: ±30 - ±600V | |||||||
| Accuracy | ±2%rdg±3dgt | |||||||
| Current measurement | ||||||||
| Measuring range | 0.00nA - 5.50mA | |||||||
| Accuracy | ±10%rdg (*2) | |||||||
| Capacitance measurement | ||||||||
| Measuring range | 5.0nF - 50.0µF | 5.0nF - 25.0µF | ||||||
| Accuracy | ±5%rdg±5dgt | |||||||
| General | ||||||||
| Power source | Rechargeable Battery (Lead-acid Battery) 12V (*3) Charging power: DC 15VA MAX | |||||||
| Communication Interface | Bluetooth: Ver2.1 + EDR Class2, USB: Ver1.1 | |||||||
| Applicable standards | IEC 61010-1, 61010-2-030 CAT Ⅳ 600V Pollution degree 2, IEC 61010-031, IEC 61326-1, 2-2 | |||||||
| Dimension | 208 (L) × 225 (W) × 130 (D) mm 380 (L) × 430 (W) × 154 (D) mm (Hard case) | |||||||
| Weight | 3127: 4kg Approx. (including battery), Total: 8kg Approx. (including Accessories) | |||||||
| Accessories | 7165A (Line probe), 7224A (Earth cord), 7225A (Guard cord), 8019 (Hook type prod), 8327EU (Power adaptor 15V/1A), 9171 (Hard case), Instruction manual | |||||||
| Optional | 7168A (Line probe with alligator clip: 3m), 7253 (Longer line probe with alligator clip: 15m), 8258 (USB communication set), 8302 (Adaptor for recorder 1mV/1µA) | |||||||